Journal Information
Vol. 35. Issue 4.
Pages 235.e1-235.e4 (April 2016)
Share
Share
Download PDF
More article options
Visits
5165
Vol. 35. Issue 4.
Pages 235.e1-235.e4 (April 2016)
Case report
Open Access
Massive hemoptysis as an unusual complication of right heart catheterization: Successful treatment with percutaneous stent
Hemoptises massivas como complicação pouco habitual do cateterismo direito: um tratamento bem-sucedido com stent percutâneo
Visits
5165
Filipa Melãoa,
Corresponding author
filipamelao@gmail.com

Corresponding author.
, Ricardo Lopesa, Rui André Rodriguesa, Domingos Magalhãesa, Paulo Vilares Morgadob, João Carlos Silvaa, Manuel Campeloa,c, Maria Júlia Maciela
a Serviço de Cardiologia, Centro Hospitalar São João, Porto, Portugal
b Serviço de Radiologia, Centro Hospitalar São João, Porto, Portugal
c Faculdade de Medicina da Universidade do Porto, Portugal
This item has received

Under a Creative Commons license
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (3)
Show moreShow less
Abstract

The authors present a case report of successful treatment of pulmonary artery perforation by endovascular stent graft implantation.

Keywords:
Swan-Ganz catheterization
Hemoptysis
Resumo

Os autores descrevem um caso de sucesso do tratamento da rotura da artéria pulmonar com a implantação de um stent endovascular.

Palavras-chave:
Cateterização Swan-Ganz
Hemoptises
Full Text
Introduction

Pulmonary artery catheterization is a useful tool for the diagnosis and management of cardiovascular and pulmonary disease.1,2 A rupture in a branch of the pulmonary artery is a very rare complication of right heart catheterization using a Swan-Ganz catheter; however, it does have a high mortality rate due to immediate and delayed bleeding following pseudoaneurysm formation.3

Case report

A 79-year-old woman with a medical history of atrial fibrillation and rheumatic mitral valve disease underwent closed mitral commissurotomy in 1972. In 2014, the patient had New York Heart Association class III symptoms and moderate mitral stenosis, with a valve area of 1.4 cm2 by planimetry, and moderate pulmonary hypertension. In order to better assess the pulmonary hypertension, and as part of a planned mitral surgery, elective right heart catheterization via the right femoral vein was performed using a 7F Swan-Ganz pulmonary artery catheter. The patient was under anticoagulation (warfarin) and the procedure was performed with bridging therapy (INR of 1.56 on the day of the procedure).

During catheterization, after inflation of the balloon tip of the pulmonary artery catheter in the wedge position, she suddenly developed massive hemoptysis and became hypoxic. She was sedated and given aminocaproic acid and vitamin K. The acute bleeding was stopped and the patient was admitted to the coronary care unit. Intubation or mechanical ventilation were not necessary.

A contrast-enhanced computed tomography scan of the chest showed a pulmonary artery pseudoaneurysm with a maximum diameter of 45 mm at the right middle lobe, which was enhanced by contrast (Figure 1A and B). Pulmonary angiography confirmed the diagnosis (Figure 2A). After consultations with the respiratory medicine and interventional radiology departments, a percutaneous stent graft implantation was planned, in which a 6F multipurpose guiding catheter (Cordis) was placed and a PT2TM guide wire was positioned distal to the aneurysm. Under angiographic control a 3.5 mm×18 mm stent (BeGraft Peripheral Stent Graft System, Bentley Innomed) was placed in the pulmonary artery. This procedure was successful and there were no complications (Figures 2B and C).

Figure 1.

Contrast-enhanced chest computed tomography scan showing the pulmonary artery pseudoaneurysm (arrows): (A) coronal view; (B) sagittal view.

(0.24MB).
Figure 2.

Pulmonary angiogram confirming the presence and location of the pulmonary artery pseudoaneurysm (arrow), showing its extent: (A) anteroposterior view, with stent graft implantation (arrow) to treat the pseudoaneurysm; (B) anteroposterior view, showing an excellent result with unrestricted blood flow into the periphery when the bleeding had stopped (arrow); (C) anteroposterior view.

(0.21MB).

During hospitalization, the patient was hemodynamically stable and had no further hemoptysis. She was discharged 16 days after admission, treated with aspirin 100 mg/day, clopidogrel 75 mg/day, warfarin 5 mg (dose set by the hematology department), furosemide 40 mg/day and pantoprazole 20 mg/day. A follow-up chest computed tomography scan performed two months later showed patency of the stent graft in the pulmonary artery, with a completely thrombotic false aneurysm (Figure 3). At seven months of follow-up, no events were recorded.

Figure 3.

Contrast-enhanced chest computed tomography scan (sagittal view) showing the stent in the pulmonary artery without contrast leak (arrow).

(0.18MB).
Discussion

Iatrogenic perforation of a pulmonary artery, followed by pseudoaneurysm formation within minutes or even months, is an uncommon but often catastrophic complication in cases of rupture.3–6 The reported incidence is between 0.01% and 0.47%, with a mortality rate of 50%–75%.1 A pseudoaneurysm contributes to recurrent hemorrhage in 30–40%.6

Risk factors for pulmonary artery rupture include female gender, age over 60 years, pulmonary hypertension, systemic anticoagulation therapy, long-term steroid use, inappropriate catheter placement, peripheral location of the pulmonary artery catheter, surgically induced hypothermia and cardiac manipulation during surgery.5,7 Our patient had at least three risk factors; the question of inappropriate positioning of the catheter is difficult to assess.

Damage to the arterial wall may be due to the catheter tip being advanced too distally, distal migration of the catheter during balloon deflation, retraction of an inflated balloon, or high-pressure inflation of the balloon in excess of the tensile strength of the vessel wall.6,8,9

Preventive measures to reduce the risk of catheter-induced injury during catheter placement are essential, and include inflating the balloon in a large proximal artery; floating the catheter to its wedge position; minimizing the time in wedge position; avoiding excessive catheter manipulation or balloon hyperinflation; inflation of the catheter balloon with air, not fluids; and deflating the balloon while traction is applied on the catheter.7,8

Catheter-induced injury often affects elderly women and occurs mainly on the right side.8,9 Hemoptysis is the hallmark symptom.8 Clinical presentation may also include hypoxemia, hypotension, bronchospasm, pleural effusion, pneumothorax and hemothorax; however, the patient may remain asymptomatic.4,5,8 Death usually occurs secondary to asphyxia.1

Primary management of this injury focuses on the prevention of asphyxia.1 A high level of suspicion, prompt diagnosis and appropriate treatment are essential to avoid mortality. Differential diagnosis includes pulmonary edema, pulmonary infarction, and trauma of the tracheobronchial tree.8

Initial measures include provision of adequate oxygenation and minimization of coughing, airway suctioning, and straining; in certain cases lung isolation with selective bronchial intubation may be necessary.1,8 Before the development of percutaneous endovascular techniques, pulmonary artery pseudoaneurysm would have been treated invasively by surgical lobectomy.3 Currently, coil embolization is commonly used as the first-choice treatment, with success in 75% of cases.4 Embolization also may be performed with n-butyl-cyanoacrylate.6 A successful case was reported using direct injection of thrombin through the distal port of the Swan-Ganz catheter during right heart catheterization.10 Stent graft implantation has been described in two previous case reports.7,8

Conclusion

Although the incidence of pulmonary artery perforation by the Swan-Ganz catheter is low, the high potential for rapid mortality makes it essential to be familiar with diagnosis and treatment. Coil embolization is commonly used as the first-choice treatment of pulmonary artery rupture and pseudoaneurysm. This case supports stent graft implantation as an alternative interventional approach for the management of a pulmonary pseudoaneurysm.

Ethical disclosuresProtection of human and animal subjects

The authors declare that no experiments were performed on humans or animals for this study.

Confidentiality of data

The authors declare that they have followed the protocols of their work center on the publication of patient data.

Right to privacy and informed consent

The authors declare that no patient data appear in this article.

Conflicts of interest

The authors have no conflicts of interest to declare.

References
[1]
M. Fortin, R. Turcotte, O. Gleeton, et al.
Catheter-induced pulmonary artery rupture: using occlusion balloon to avoid lung isolation.
J Cardiothorac Vasc Anesth, 20 (2006), pp. 376-378
[2]
A. Markwick, R. Butler.
Pulmonary artery catheterization.
Medicine, 42 (2014), pp. 475-476
[3]
A. Gallardo, J. Jiménez-Mazuecos, A. Gutiérrez, et al.
Massive hemoptysis as an unusual complication of right heart catheterization. Temporary hemostasis with a Berman diagnostic catheter.
Rev Esp Cardiol, 61 (2008), pp. 99-101
[4]
M. Asano, G. Gäbel, O. Allham, et al.
Pulmonary artery pseudoaneurysm in a patient with aortic valve stenosis.
Ann Vasc Surg, 27 (2013), pp. 238
[5]
M.C. Schechter, E. Prezzi, G. Cabral, et al.
Ruptura da artéria pulmonar pelo cateter de Swan-Ganz. Relato de caso.
Rev Bras Ter Intensiva, 21 (2009), pp. 226-230
[6]
A. Zuffi, G. Biondi-Zoccai, F. Colombo.
Swan-Ganz-induced pulmonary artery rupture: management with stent graft implantation.
Catheter Cardiovasc Interv, 76 (2010), pp. 578-581
[7]
S. Keymel, M.W. Merx, T. Zeus, et al.
Stenting as a rescue treatment of a pulmonary artery false aneurysm caused by Swan-Ganz catheterization.
Case Rep Pulmonol, (2014),
[8]
B.J. Muller, A. Gallucci.
Pulmonary artery catheter induced pulmonary artery rupture in patients undergoing cardiac surgery.
Can Anaesth Soc J, 32 (1985), pp. 258-264
[9]
M.H. Mullerworth, P. Angelopoulos, M.A. Couyant, et al.
Recognition and management of catheter-induced pulmonary artery rupture.
Ann Thorac Surg, 66 (1998), pp. 1242-1245
[10]
D.R. Dobies, A. Cohoon, A. Bates.
Repair of a perforated pulmonary artery due to a Swan-Ganz catheter using thrombin injection.
Circulation, 119 (2009), pp. e521-e522
Idiomas
Revista Portuguesa de Cardiologia (English edition)
Article options
Tools
en pt

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos

By checking that you are a health professional, you are stating that you are aware and accept that the Portuguese Journal of Cardiology (RPC) is the Data Controller that processes the personal information of users of its website, with its registered office at Campo Grande, n.º 28, 13.º, 1700-093 Lisbon, telephone 217 970 685 and 217 817 630, fax 217 931 095, and email revista@spc.pt. I declare for all purposes that the information provided herein is accurate and correct.