Journal Information
Vol. 34. Issue 3.
Pages 211.e1-211.e3 (March 2015)
Share
Share
Download PDF
More article options
Visits
4467
Vol. 34. Issue 3.
Pages 211.e1-211.e3 (March 2015)
Case report
Open Access
When inappropriate becomes beneficial
Quando o inapropriado se torna saudável
Visits
4467
José David Arroja
Corresponding author
jose-david.arroja@latour.ch

Corresponding author.
, Marc Zimmermann
Cardiovascular Department, Hôpital de La Tour, Meyrin-Geneva, Switzerland
This item has received

Under a Creative Commons license
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (1)
Abstract

We report the case of a young man who accidentally received a prolonged electric discharge from electrical wires and released the electric source with the help of an inappropriate shock from his implantable cardioverter-defibrillator (ICD), after misinterpretation of the electrical signal by the device as a ventricular tachycardia. This case illustrates the “electrical noise” phenomenon, and underscores the need for precautions for patients with an ICD and their physicians.

Keywords:
Implantable cardioverter-defibrillator
Inappropriate shock
Electrocardiographic noise
Resumo

Relatamos o caso de um jovem que acidentalmente recebeu uma descarga eléctrica prolongada por contacto com fios eletrónicos. Foi ajudado a libertar a fonte elétrica pela administração inapropriada de um choque pelo cardioversor desfibrilador implantável (CDI), depois de uma interpretação errada do sinal elétrico pelo dispositivo, lido como uma taquicardia ventricular. Este caso ilustra o fenómeno de «ruído elétrico», e sublinha as precauções necessárias para os doentes portadores de CDI e para os seus médicos.

Palavras-chave:
Cardioversor Desfibrilador Implantável
Choque inapropriado
Ruído eletrocardiográfico
Full Text
Case report

We report the case of a 36-year-old male with arrhythmogenic right ventricular cardiomyopathy and an implantable cardioverter-defibrillator (ICD) since 2007 for primary prevention of sudden cardiac death. While abroad on holiday, the patient received an electric discharge while manipulating computer wires. On his return two weeks later, he made an appointment at our pacemaker clinic in order to confirm the proper functioning of his device. The patient had not sought medical assistance in the meantime, and denied any cutaneous or other lesion related to the event.

ICD interrogation revealed the application of a shock 11 seconds after the beginning of continuous high-frequency electrical activity, interpreted as an episode of ventricular tachyarrhythmia by the device.

The review of the ICD electrogram (Figure 1) clearly shows atrioventricular activity with narrow and regular QRS complexes parasitized by high-frequency low-intensity baseline interference, corresponding to the accidental 50–60 Hz electrical exposure, terminated by the ICD shock delivery. Unlike the case reported by Saguner and Duru1 in which a shock was avoided by appropriate detection of interference by the ICD's noise reversion algorithm, our patient's device was not equipped with this functionality, and hence was unable to respond adequately, delivering a 35-J shock that probably helped the patient release the electrical source, as noise termination occurred immediately after shock delivery and as the patient himself reported.

Figure 1.

Implantable cardioverter-defibrillator (ICD) electrogram with (1) start of electrical interference and (2) ICD shock delivery and termination of interference.

(2.47MB).

This case highlights a rare situation in which an inappropriate shock can be beneficial and underscores the importance of advising patients carrying a defibrillator to avoid any electrical exposure.2,3

Ethical disclosuresProtection of human and animal subjects

The authors declare that no experiments were performed on humans or animals for this study.

Confidentiality of data

The authors declare that no patient data appear in this article.

Right to privacy and informed consent

The authors declare that no patient data appear in this article.

Conflicts of interest

The authors have no conflicts of interest to declare.

References
[1]
A.M. Saguner, F. Duru.
Implantable cardioverter defibrillator avoids shock during electrocution.
Eur Heart J, 33 (2012), pp. 694
[2]
E. Occhetta, M. Bortnik, A. Magnani, et al.
Inappropriate implantable cardioverter-defibrillator discharges unrelated to supraventricular tachyarrhythmias.
Europace, 8 (2006), pp. 863-869
[3]
D. Graf, E. Pruvot.
Inappropriate ICD shocks.
Copyright © 2014. Sociedade Portuguesa de Cardiologia
Idiomas
Revista Portuguesa de Cardiologia (English edition)
Article options
Tools
en pt

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos

By checking that you are a health professional, you are stating that you are aware and accept that the Portuguese Journal of Cardiology (RPC) is the Data Controller that processes the personal information of users of its website, with its registered office at Campo Grande, n.º 28, 13.º, 1700-093 Lisbon, telephone 217 970 685 and 217 817 630, fax 217 931 095, and email revista@spc.pt. I declare for all purposes that the information provided herein is accurate and correct.